top of page

Efficient Tips for Column Cleaning in Reverse Phase Liquid Chromatography




We will explore straightforward yet highly effective tips for column cleaning, particularly geared toward reverse-phase liquid chromatography.


#1. Follow Manufacturer's Recommendations It's crucial to adhere to the manufacturer's recommended cleaning procedure for your column.


#2. Use Cleaning Solvents: Using 10 to 20-column volumes of cleaning solvents during the cleaning process. Here's a suggested sequence for generic reverse-phase columns when manufacturer guidelines are unavailable:


a) Water: Start with water to eliminate polar impurities or polarized compounds.


b) Methanol: Follow up with methanol.


c) Acetonitrile (ACN): Use ACN as a miscible organic solvent, especially when using heptane.


d) Heptane: Employ heptane for removing highly retained non-polar impurities.


e) Isopropanol (IPA): Use IPA to bridge the gap between ACN and heptane due to its miscibility with both.


f) ACN (Again): Continue with ACN.


g) ACN with 0.1% Formic Acid: For thorough cleansing, employ an acidic mobile phase.


h) Methanol (Again): Follow up with methanol.


#3. Store in Organic Solvent to preserve your reverse phase liquid chromatography column, store it in a solution of 20% water and 80% methanol, especially when used for protein analysis.


#4. Revive Blocked Flow: If the flow is obstructed due to extended column use, consider using a 5 to 10 molar urea solution to denature proteins deposited on the column. Restore flow by washing with urea, followed by acidified methanol-water or acid water (acetic or formic acid to acidify the mobile phase).


#5. EDTA for Metal Ion Contamination: In cases of suspected metal ion contamination, use EDTA as an effective cleansing agent. EDTA can form complexes with metal ions, removing them from the column. Afterward, rinse the column thoroughly with water to eliminate EDTA.


#6. Reverse Flow Direction: During column cleaning, reverse the flow direction to prevent contaminants from traversing the entire column length. Typically, contaminants accumulate near the inlet, so reversing the flow from the outlet side reduces the distance they need to travel.


#7. Manufacturer's Guidelines and Adaptation: Always consider the manufacturer's directions, as the number of column volumes and cleaning times may vary based on contamination.


In summary, these simple yet effective column-cleaning tips will help maintain the performance and longevity of your reverse-phase liquid chromatography column.


bottom of page